This high-frequency infrared carbon sulfur analyzer is designed specifically for metal material testing. The high-frequency furnace utilizes international military-grade, high-precision ceramic capacitors to adapt to the combustion characteristics of samples with varying melting points, ensuring high and stable conversion rates. The infrared detection system utilizes non-dispersive infrared absorption, coupled with spatial temperature control technology, resulting in the instrument's core technologies, including long-term stability, repeatability, and accuracy, leading the world.
Modular and Integrated
The high-frequency infrared carbon sulfur analyzer is modular and integrated design, stable and reliable
Multi-point Correction
Full range calibration technology, single point and multi-point correction
Third Generation High Frequency Furnace Head
Using the third generation high frequency furnace head, the infrared carbon and sulfur analyzer integrates the internal heating of the furnace head, dust reverse blowing, dust suction and air intake, dust removal brush into one, with complete functions and high integration;
Oscilloscope
The function of oscilloscope can quickly diagnose the reliability of instrument circuit and improve the efficiency of instrument use
Self-inspection
The system of the high-frequency infrared carbon sulfur analyzer has self-inspection function, and the software controls multiple automatic leak detection, which can effectively determine the gas tightness of the instrument
Remote Diagnosis
Remote diagnosis function, can realize online remote control of instruments and diagnosis
Automatic Alarm and Protection
Automatic alarm and protection function, when personnel operation error or instrument work is not normal, the instrument stops by itself;
Patented Technology
This high-frequency infrared carbon sulfur analyzer has the patented technology of high frequency radiation shielding to reduce the harm of electromagnetic radiation.
| MODEL | SES-802 | |
| ANALYZE THE VARIETY | Blackmetals, non-ferrous metals, alloy materials, some ores, titanium dioxide, new energy materials, and other non-metallic materials | |
| ANALYSIS SCOPE | Low carbon: 0.0001-0.5% High carbon:0.05-99.99% | (Any combination of four measuring pools) |
| Low sulfur: 0.0001-0.5% High sufur:0.05-99.99% | ||
| SENSITIVITY | Carbon:0.1ppm Sulfur 0.1ppm | |
| ACCURACY | Carbon: 1ppm or RSD≤0.8% (one of the two conditions is met) | |
| REFERENCE STANDARDS | Carbon meets 1S09556 standard, sulfur meetsI504935 standard Complies with JG395-2016 "Measurement Regulations for Carbon and Sulfur Analyzer" GB/T 20123-2006/1S15350.2000 "Determination of Total Carbon and Sulfur Content in lron and Steel-Infrared Absorption Method after High Frequency lnduction Furnace Combustion" | |
| DETECTION | Non dispersive infrared absorption method, high-sensitivity carbon infrared detection | |
| ANALYSIS TIME | Adjustable within 45 seconds, the analysis endpoint can be controlled based on both absorbance and time | |
| ANALYTICAL ABILITY | Usually 60 samples per hour | |
| CORRECTION METHOD | Single point, multi-point, linear correction, can establish first, second, third, fourth, and fifth correction cuves | |
| POWER SUPPLY | 220V/50Hz±5% | |
| HIGH FREQUENCY FURNACE POWER | 2.7KW/20MHz, dual air cooling system | |
| HEATING TEMPERATURE | 2000′C (infinitely adjustable) | |
| DUST REMOVAL METHOD | Intellgent high-pressure blowback combined with fixed frequency dust brushing | |
| SMART STAY | Equipped with intelligent sleep function,the instrument automatically shuts off power and gas when there is no operation | |
| DATA INTERFACE | USB, RS232, Free choice of network card, etc. | |
| REQUIRED GAS | Oxygen, 99.5% purity, 35psi (2.41bar)+10%. | |
| EXTERNAL DIMENSIONS | L550*W695^H867mm | |
| STANDARD SAMPLE | C STANDARD VALUE(%) | S STANDARD VALUE(%) | C TEST VALUE(%) | S TEST VALUE(%) |
| carbon structural steel YSBC111356-2011 | 0.338 | 0.014 | 0.3382 | 0.0140 |
| 0.3386 | 0.0141 | |||
| 0.3377 | 0.0141 | |||
| 0.3381 | 0.0142 | |||
| 0.3379 | 0.0139 |