This high-frequency infrared carbon sulfur analyzer is a newly upgraded product from Science, targeting the high-end market. The infrared carbon sulfur analyzer utilizes a third-generation high-frequency furnace system and a double-layer infrared isolation system. The high-frequency furnace utilizes internationally leading vacuum ceramic capacitor oscillation technology. The high-power high-frequency combustion furnace provides the instrument with excellent sample combustion performance. The infrared carbon and sulfur analyzer also features adjustable power, matching the power to the sample analysis method. Once set, lifelong use ensures sufficient combustion of different samples while allowing the instrument to operate at less than full load. This improves sample adaptability and instrument reliability. The infrared constant temperature system and mirror-finished stainless steel piping throughouhet the instrument ensure excellent equipment repeatability and long-term stability.
High-precision vacuum ceramic capacitors
Fully mirrored stainless steel piping
Pre-purification device
Physical dehydration device
Catalytic furnace
Automatic sample injection
MODEL | SES-902 |
ANALYZE THE VARIETY | Black metals,non-ferrous metals,alloy materials,rocks and minerals,soil, titanium dioxide,newenergy materials,and othe non-metallic materials |
ANALYSIS SCOPE | Low carbon:0.00001-0.5% High carbon:0.05-99.99% Lowsulfur:0.00001-0.5% High sulfur:0.05-99.9% |
SENSITIVITY | Carbon:0.01ppm System:0.01ppm |
ACCURACY | Carbon:1ppm orRSD≤0.5%(one ofthe two conditions ismet) Sulfur.1ppmor RSD≤0.5%(one ofthe two conditions ismet) |
REFERENCE STANDARDS | Carbon meets S09556 standard,sulfur meets IS04935standard Complies with JG395-2016"Measurement Regulationsfor Carbon and Sulfur Analyzer'GB/T20123-2006/S15350.2000 "DeterminationofTotal Carbon and Sulfur Content in lron and Steel-Inrared Absorption Method after High Frequency lnduction Fumace Combustion° |
DETECTION | Non dispersive infraredabsorption method,high-sensitivitycarbon infrared cell |
ANALYSIS TIME | Adjustable within 45seconds,the analysisendpoint can be controlled based on both absorbance and time |
ANALYTICALABILITY | Usually 60 samplesper hour |
CORRECTION METHOD | Single point,multi-point,linear corection,canestablish fist,second, third,fourth,and fith correction curves |
POWERSUPPLY | 220V/50Hz+5% |
HIGH FREQUENCYFURNACE POWER | 3.5KW/20MHz,dualaircolingsystem,quickloading andunloading componensforcombustion tubes |
ADJUSTABLE POWER | Equippedwith power regulation function,automaticalymatched with analysismethods |
HEATING TEMPERATURE | 2000℃(infinitely adjustable) |
DUSTREMOVAL METHOD | Intelient dualzonefully automatic dust removal,integratinghigh-pressure blowback,dust brushing,slagand dust separation,dust suction,and dust collection inone |
SMART STAY | Equipped withintelligent sleep function,the instrument automatically shuts off powerand gaswhen there is no operation |
DATAINTERFACE | USB、R5232、Free choice of netwok card,etc |
REQUIRED GAS | Gas purity of99.5%6,35psi(2.41bar'+10%. |
INFRARED CONSTANTTEMPERATURE | Equippedwith an automatic temperature controlsystem,with a temperature iference of±0.5℃ |
ANALYZE PARAMETERS | Flexible changes can be made according to sample requirements,and standardcurves can beestablished separately foreach analysis condition |
GAS PIPELINE | ullmirorstainless steel pipeline with strong corrosion resistance and long-tem airtightnes5 |
CAPACITORTYPE | High precision vacuum ceramic capacitor |
OPTIONAL DEVICES | Pre purification device,water removal device,catalyticfurmace,dual gas path |
EXTERNAL DIMENSIONS | L600*W704"H870mm |
STANDARD SAMPLE | C STANDARD VALUE(%) | S STANDARD VALUE(%) | C TEST VALUE (%) | S TEST VALUE (%) |
Silicomanganese alloy YSBC37651B-10 | 2.26 | 0.04 | 2.2608 | 0.0403 |
2.2601 | 0.0401 | |||
2.2604 | 0.0405 | |||
2.2611 | 0.0398 | |||
2.2606 | 0.0402 | |||
2.2603 | 0.0399 |